| Model Number |
LifeVest WCD 4000 System |
| Medical Device Problem Codes |
Charging Problem (2892); Contamination /Decontamination Problem (2895)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/03/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A us distributor reported that a charger failed due to a functional issue.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation of integrated charger has been completed.The reported problem (fails due to functional issue) was confirmed due to corrosion on the battery board.The charger displayed a light when no battery was inserted, indicating a charger failure.The root cause for the corrosion was ingress of an unknown contaminant.Lifevest patient training materials have been updated as a reminder not to expose the lifevest electronic components to liquids.There was no adverse event that resulted from the damaged charger.
|
| |
|