Medline Industries, LP MEDLINE; METER,GLUCOSE,EVENCARE G3,PROF.USE ONLY Back to Search Results
Catalog Number MPH3540
Medical Device Problem Code Low Readings (2460)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/02/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the meter is not passing low solution testing.The high solution qc test is passing just fine; only the low qc test is failing consistently.
 
Additional Manufacturer Narrative
It was reported that the item was not passing low solution, and the customer reported that the meter is not passing low solution testing while the high solution qc test is passing just fine and only the low qc test is failing consistently.Medline received an evencare g3 blood glucose monitor level 1 control solution; visual inspection found nothing out of place, the expiration date was still good as of 1/11/2024, and the color of solution and overall appearance of the bottle itself was fine.Upon further testing, medline compared new level 1 and level 3 control solutions; the new controlled solutions both met the criteria, and when testing the level 1 solution that was sent in, it showed lower numbers then needed to meet the criteria.The complaint was confirmed, a cause code of "comp./part defective, broke" was recorded, and no further action was taken because objectionable trends were not seen for this issue.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
METER,GLUCOSE,EVENCARE G3,PROF.USE ONLY
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
herchelle teodoro
three lakes drive
northfield, IL 60093--275
MDR Report Key50883462
Report Number1417592-2026-49832
Device Sequence Number10667595
Product Code CGA
UDI-Device Identifier10884389831199
UDI-Public10884389831199
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K120448
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMPH3540
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer01/10/2024
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/02/2024
Initial Report FDA Received Date09/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-