| Catalog Number |
PVE35A |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during an unknown surgery, the jaws did not open after firing.The manual override was used to return the knife and the jaws were released.There were no issues with the blood vessels.The cartridge color was white.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 9/24/2026.D4: batch #911d44.Investigation summary:
the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one pve35a device was returned with no apparent damage and with a reload loaded on the device.The reload was received partially fired 1/10.The returned device and reload were tested for functionality in the straight position.The reload was reset and reloading it into the device.The device achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the remaining staples met the staple form release criteria.It is possible that while loading the reload, the reload was pushed farther back than the reload alignment slot resulting in the knife pushing the one-piece sled forward and locking the reload.The event described of difficult to open could not be confirmed as the device opened and closed without any difficulties noted.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 911d44, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|