| Catalog Number |
MX9341L |
| Medical Device Problem Codes |
Break (1069); Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/08/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, upon evaluation of the additionally received medex stopcocks, cracking was observed along the a-port shafts of the returned device.The reported defect was confirmed.There was patient involvement and moderate patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Initial reporter phone number with extension (b)(6).One sample was returned for evaluation.The lot history was reviewed, and no nonconformities were identified that could have contributed to the reported complaint.Visual inspection and functional testing were performed on the returned device.The device exhibited cracking along the a-port shaft, confirming the reported defect.The location and nature of the cracking are consistent with stresses that may result from excessive applied torque, such as over-tightening during connection.Based on the evaluation findings, the observed breakage is consistent with loading conditions exceeding those expected during normal use.
|
| |
|