(b)(4)
date sent: 9/24/2026
d4: batch #991d21.Investigation summary:
the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one ec60a device was returned with the jaws and closure trigger in the closed position, with no apparent damage and with no reload present.Also, the knife was noted partially advanced, the red manual knife reverse button was activated and the knife returned to the home position as expected.The device was tested for functionality in the articulated position with a test reload and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples met the staple release criteria.It should be noted that when using the reverse button ensure that the knife is fully back on its home position, if the knife remains advanced the device jaws would not open.The event described could not be confirmed as the device performed without any difficulties noted.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 991d21, and no non-conformances were identified.A manufacturing record evaluation was performed for the finished device ec60a lot number a9m93p and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|