• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Wilson-Cook Medical Inc INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICE; Hemostatic Clip

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Wilson-Cook Medical Inc INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICE; Hemostatic Clip Back to Search Results
Model Number G58010
Medical Device Problem Codes Separation Failure (2547); Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/27/2026
Type of Reportable Event Malfunction
Event or Problem Description
During an endoscopic clipping procedure, the physician used a cook instinct plus endoscopic clipping device.It was reported that while the clip was being inserted through the scope, the staff member was holding the red part of the handle, the clip could not be deployed.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.A photo of the device coiled in the pouch was provided, the lot number of the label matches that in the report.The device history record for the lot number said to be involved was reviewed.Nonconformances that could potentially be related to the complaint were contained in the associated dhr.The nonconformance documentation supports the affected device(s) were dispositioned appropriately prior to release of this lot.There is no evidence nonconforming product was released for distribution.In an effort to heighten awareness of the potential connection of the customers report to the current manufacturing processes production personnel were notified.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.The instructions for use states, "note: if separation of the clip is not immediate, gently move the catheter back and forth or use other endoscopic maneuvers to separate the catheter from the clip." prior to distribution, all instinct plus endoscopic clipping devices are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: based on the quality engineering risk assessment no corrective action is warranted at this time.A review of the complaint history was conducted.Based on this review, the likelihood of this type of report is rare.Quality assurance will continue to monitor for complaint trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICE
Common Device Name
Hemostatic Clip
Manufacturer (Section D)
Wilson-Cook Medical Inc
4900 bethania station rd
winston salem NC 27105
Manufacturer (Section G)
Wilson-Cook Medical Inc
4900 bethania station rd
winston salem NC 27105
Manufacturer Contact
sabrina o¿brien
4900 bethania station rd
winston salem, NC 27105
+1(336)744-0157
MDR Report Key50883693
Report Number1037905-2026-00583
Device Sequence Number13441215
Product Code PKL
UDI-Device Identifier00827002580107
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K212323
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG58010
Device Catalogue NumberINSC-P-7-230-S
Device Lot NumberW5029884
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/27/2026
Initial Report FDA Received Date09/24/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/06/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Olympus Scope, unknown model.
Patient SexUnknown
-
-