| Model Number |
D233 |
| Medical Device Problem Codes |
Pacing Problem (1439); Inappropriate/Inadequate Shock/Stimulation (1574)
|
| Health Effect - Clinical Code |
Shock from Patient Lead(s) (3162)
|
| Date of Event |
08/27/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) delivered an inappropriate burst of anti-tachycardia pacing (atp) and three inappropriate shocks.Technical services (ts) was consulted and agreed with the local field representative that the shocks were delivered for supraventricular tachycardia.Ts discussed proposed changes to the rate cutoff and discriminator changes in the ventricular tachycardia monitor only zone with the physician.No additional adverse patient effects were reported.This icd remains in service.
|
| |
|