• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Black Diamond Video, Inc. HexaVue; Custom Room Rack

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Black Diamond Video, Inc. HexaVue; Custom Room Rack Back to Search Results
Model Number HexaVue SCC 1.0
Medical Device Problem Code Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/08/2026
Type of Reportable Event Malfunction
Event or Problem Description
The user facility reported that a monitor and wall display connected to their hexavue custom room rack were ¿flickering¿.No report of procedure involvement.No report of injury.
 
Additional Manufacturer Narrative
A steris service technician arrived onsite following the reported event to inspect the hexavue custom room rack.During the technician¿s inspection, he was unable to duplicate the reported event and found no issues with the function or operation of the hexavue custom room rack, monitor and wall display, the devices were operating to specification.The technician returned the hexavue custom room rack to service and no additional issues have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HexaVue
Common Device Name
Custom Room Rack
Manufacturer (Section D)
Black Diamond Video, Inc.
503 canal blvd
richmond CA 94804
Manufacturer (Section G)
Black Diamond Video, Inc.
503 canal blvd
richmond CA 94804
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
+1(440)259-7401
MDR Report Key50883723
Report Number3008776287-2026-00692
Device Sequence Number10667602
Product Code KQM
UDI-Device Identifier00724995198503
UDI-Public(01)00724995198503(10)2.1.0
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberHexaVue SCC 1.0
Device Catalogue NumberXQHILCC1
Device Lot Number(10)2.1.0
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/08/2026
Initial Report FDA Received Date09/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/10/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-