| Model Number |
106531US |
| Medical Device Problem Code |
Failure to Charge (1085)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the log files were submitted for evaluation for backup battery fault alarms.The backup battery was changed in the primary system controller and backup battery fault alarms persisted despite syncing clock, they waited 10 minutes after changing, and there was 28 months left on the new battery.It appeared the event log file recorded a backup battery fault event on 02sep2026 at 2:27:37.This event appeared to have occurred after the system controller attempted a load test.The primary system controller was exchanged and the alarms resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
No further information was provided.A supplemental report will be reported once the manufacturer¿s investigation is completed.
|
| |
|