The customer reported that (b)(4) custom packs had the sterile water "explode" within the pack contaminating everything contained inside the pack; there was no patient contact.A photo and a sample were received, current stock was reviewed, finished good production records were reviewed, and the finished good pack build was reviewed; upon further inspection, the packs were soaked with water and the seals of the water were broken.The reported issue was confirmed based on the sample evaluation.No defects were found in the manufacturing facility, the pack's component locations likely did not contribute to the reported issue, and medline was unable to confirm that it was frozen during shipping, so this was considered an unknown responsible party mishandling issue as any pressure on the bottle could cause this issue; the supplier quality team will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|