Catalog Number VAL011CTASD
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/31/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that 23 of their custom packs had the sterile water "explode" within the pack contaminating everything contained inside the pack.
 
Additional Manufacturer Narrative
The customer reported that (b)(4) custom packs had the sterile water "explode" within the pack contaminating everything contained inside the pack; there was no patient contact.A photo and a sample were received, current stock was reviewed, finished good production records were reviewed, and the finished good pack build was reviewed; upon further inspection, the packs were soaked with water and the seals of the water were broken.The reported issue was confirmed based on the sample evaluation.No defects were found in the manufacturing facility, the pack's component locations likely did not contribute to the reported issue, and medline was unable to confirm that it was frozen during shipping, so this was considered an unknown responsible party mishandling issue as any pressure on the bottle could cause this issue; the supplier quality team will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
DBD-CATARACT PACK
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
sasidhar lakku
three lakes drive
northfield, IL 60093--275
MDR Report Key50883823
Report Number1417592-2026-49890
Device Sequence Number9939817
Product Code OJK
UDI-Device Identifier40195327152032
UDI-Public40195327152032
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberVAL011CTASD
Device Lot Number22LBH455
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer02/08/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/31/2023
Initial Report FDA Received Date09/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Patient Sequence Number1