It was reported that the cracked light handle cover.Medline's investigation received one photo and a physical sample for one light handle cover; the rest of the pack was not available, finished good production records were reviewed, and finished good pack build was reviewed.Medline observed the light handle cover out of its original packaging with a crack on the handle, confirming the reported issue.Further investigation reveals that the root cause is attributed to a vendor product issue related to the component manufacturing process; this complaint has been provided to the vendor for further evaluation and supplier quality will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|