| Model Number |
PCF-H190I |
| Medical Device Problem Code |
Device Reprocessing Problem (1091)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer returned the colonovideoscope for a separate issue.During the initial evaluation of the returned device, foreign objects were observed in the air/water cylinder, air/water tube, and c-cover.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of event is unknown.Additional information will be provided if available.Based on the final investigation of the device the initial evaluation findings were confirmed.The root cause for the reportable malfunctions identified during the evaluation could not be established.Olympus will continue to monitor the field performance of this device.Should any additional relevant information become available, a supplemental report will be submitted.
|
| |
|