The customer reported the ultrasound gastrovideoscope had a separate issue.During the initial evaluation of the device, it was observed that the adhesive around objective lens was peeled, acoustic lens was damaged (deep cut), ultrasonic connector was sticky, and the ultrasound image was not displayed.No patient involvement.
|
|
The device was returned to olympus for inspection.Based on the final investigation of device and the initial evaluation findings, adhesive around objective lens was peeled, acoustic lens was damaged (deep cut), ultrasonic connector was sticky, and the ultrasound image was not displayed were confirmed.A root cause could not be identified.Based on the results of the investigation, the most probable cause of the malfunctions, 'adhesive around objective lens was peeled, acoustic lens was damaged (deep cut), and the ultrasound image was not displayed' is traced to component failure, and the cause of 'ultrasonic connector was sticky' could not be established.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|