• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; ASPIRATOR,SUCTION,W/850CC CANISTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; ASPIRATOR,SUCTION,W/850CC CANISTER Back to Search Results
Catalog Number HCS7001
Medical Device Problem Code Suction Failure (4039)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/30/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "item is not suctioning.".
 
Additional Manufacturer Narrative
The customer reported ¿no suction.¿ medline quality received one sample, visually inspected the sample and it appeared to be in good condition; the sample was plugged into a power source and turned on, and the sample turned on but produced no suction.As a result, medline was able to confirm the complaint.The cause code was updated to unknown responsible party - cannot determine cause with received sample, and the division will monitor for trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
ASPIRATOR,SUCTION,W/850CC CANISTER
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
kumar vigrahala
three lakes drive
northfield, IL 60093--275
MDR Report Key50883977
Report Number1417592-2026-49917
Device Sequence Number21499638
Product Code BTA
UDI-Device Identifier10193489036381
UDI-Public10193489036381
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K203135
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberHCS7001
Device Lot Number6462109
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer11/30/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/30/2023
Initial Report FDA Received Date09/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-