| Model Number |
DDPA2D4 |
| Medical Device Problem Codes |
Pacing Problem (1439); Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the right ventricular (rv) lead exhibited ventricular undersensing resulting in inappropriate pacing. the rv lead remains in use. it was also reported that the patient received possible inappropriate anti-tachycardia pacing (atp) for a ventricular fibrillation (vf) episode that was detected as supra ventricular tachycardia (svt). the implantable cardioverter defibrillator (icd) remains in use. no patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|