| Model Number |
MX40 - 2.4 |
| Medical Device Problem Code |
Defective Alarm (1014)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This report is based on information provided by a philips remote service personnel and has been investigated by the philips complaint handling team.Philips received a complaint on the mx40 2.4 ghz smart hopping indicating that when a patient had a 27-beat run of supraventricular tachycardia (svt), the monitor did not alarm, so the arrythmia was not noticed right away.The default configuration had svt arrythmia alarms turned off; however, it was unclear if the svt alarm was turned off or on at the time of this event.The default configuration for an svt alarm is a heartrate of 180 and a run of 10.When the patient had the arrythmia event, the patient's heartrate was in the 130's.There was no report of actual harm.
|
| |
|
Additional Manufacturer Narrative
|
|
A philips clinical transformation manager spoke to the customer, who confirmed that they lowered the hr parameter for the svt alarm to resolve the issue.The device was confirmed to be operating per specifications and no failure was identified.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
|
| |
|