Respironics, Inc. Respironics; Ventilator, Continuous, Minimal Ventilatory Support, Facility Use
|
Back to Search Results |
|
| Model Number |
V60 V60Plus Ventilator |
| Medical Device Problem Code |
Inadequate User Interface (2958)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/15/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Philips received a complaint by a biomedical engineer (bme) on the v60 indicating a device malfunction as touchscreen calibration failed.There was no patient involvement at the time the issue was discovered as the issue occurred during preventive maintenance (pm).No patient or user harm was reported.Based on the case information provided, the device did not meet product specifications.The alleged malfunction impacted the user¿s ability to use the device properly.A biomedical engineer (bme) discovered that touchscreen calibration failed during preventive maintenance (pm) and found that the replacement touchscreen needed to be ordered.Upon receiving the replacement touchscreen, the bme replaced the defective touchscreen and performed the extended self-test (est), after which the issue was resolved.The touchscreen replacement indicates that the cause of the malfunction was ultimately due to a faulty touchscreen replacement that needed to be replaced for repair.The device passed the required performance verification tests per philips standard and was returned to service.No other malfunction was found.The investigation concludes that no further action is required at this time.If additional information is received, the complaint file will be reopened.
|
| |
|