• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; SET,IRRIGATION,CYSTO,Y-TYPE,90

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; SET,IRRIGATION,CYSTO,Y-TYPE,90 Back to Search Results
Catalog Number DYND19122
Medical Device Problem Codes Disconnection (1171); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "the single-end connection tube is too flexible and kinks and pops off.".
 
Additional Manufacturer Narrative
The customer reported that the single-end connection tube is too flexible and kinks and pops off; no patient harm or medical intervention was reported.Medline quality received the reported product in its original packaging, unused condition, and visual assessment of the provided product revealed that the tube were kinked upon opening from the original packaging.The reported issue of kinked tubes was verified by visual assessment of the provided samples, and the tubing issue was the result of a compromised batch of raw material from their supplier.There are currently no known trends, and medline quality will continue to monitor for the issue through weekly complaint meetings and monthly trending analysis.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
SET,IRRIGATION,CYSTO,Y-TYPE,90
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
pranayreddy alluri
three lakes drive
northfield, IL 60093
MDR Report Key50884071
Report Number1417592-2026-49940
Device Sequence Number9939823
Product Code LJH
UDI-Device Identifier40080196993827
UDI-Public40080196993827
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K970946
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberDYND19122
Device Lot Number18322020001
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/14/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/06/2023
Initial Report FDA Received Date09/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-