| Model Number |
L321 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/17/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this pacemaker recorded a code 1003, indicating the voltage was too low for the projected remaining capacity.A request was made to have data from this device analyzed.Data analysis identified potential high impedance within the battery.Continued monitoring until radio frequency (rf) telemetry disabled was recommended.The device remains in service.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.
|
| |
|
Search Alerts/Recalls
|