It was reported that, during a thoracoscopic right lung partial resection, the clip could not be loaded into an applier properly and something was caught.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
|
|
(b)(4)
date sent 9/24/2026
d4 batch # a99j1a.Investigation summary
the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the el5ml device was returned with no apparent damage.In addition, the tyvek packaging was returned along with the instrument.In an attempt to replicate the reported incident, the device was tested for functionality.During testing, the instrument was cycled; however, on subsequent cycles, the clips failed to advance into the jaws.To further evaluate the condition of the internal components, the device was disassembled.Upon disassembly, the feeder shoe tab was found to be damaged, which prevented proper clip advancement and resulted in the feeding issue.Additionally, three (3) clips were found within the clip track.The observed feeder shoe tab damage is consistent with the reported event and confirms the failure to feed clips into the jaws.As part of ees quality process all devices are manufactured, inspected, and released to approved specifications.No conclusion could be reached regarding what may have caused the reported incident.A manufacturing record evaluation was performed for the finished device batch a99j1a number, and no non-conformances were identified.
|