| Model Number |
1944/52 |
| Medical Device Problem Codes |
Fracture (1260); Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During remote follow-up, noise which resulted in oversensing was observed on the atrial lead.The suspected cause of the event is a lead fracture; however this has not been confirmed.No intervention has been performed at this time.The patient was stable and will continue to be monitored.
|
| |
|
Event or Problem Description
|
|
Additional information received indicates that the noise was successfully reproduced with provocative testing.Diagnostic imaging was also performed to assess alleged lead damage on the rv and a leads, but the results are unavailable.
|
| |
|