| Catalog Number |
122132048 |
| Medical Device Problem Codes |
Device Dislodged or Dislocated (2923); Naturally Worn (2988)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
09/17/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that altrx neut 32idx48od was associated with liner disassociation identified on x-ray during a planned revision arthroplasty.Intra-operatively, the liner was described as worn and warped, and the ceramic femoral head exhibited substantial wear.The acetabular cup appeared intact, and the screw was flush.Patient harm was reported, and revision surgery was required; as a result, the head and liner were revised, and a new lipped liner and ts head were implanted.The implant products involved were explanted; the altrx neut 32idx48od and unk hip femoral head ceramic were discarded, and the unk hip acetabular cup pinnacle was retained by the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4)
this report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|