The customer reported that, when using the contrast injector at 200 psi of pressure, the luer lock tip of the manifold broke off.No sample or photo was received for the investigation; finished good production records and finished good pack build were reviewed, and no line-down situations, run without approvals, deviations, non-conformance reports, bom changes, or placement changes were identified that might have contributed to the reported defect.At this time, medline was unable to confirm the reported defect, as no physical sample or detailed photographic evidence was provided to support the investigation.A root cause was not determined; the cause was listed as sample required to determine cause, and the division will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|