• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; THERMOMETER,DIGITAL,ORAL,F/C,30 SEC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; THERMOMETER,DIGITAL,ORAL,F/C,30 SEC Back to Search Results
Catalog Number MDS9950H
Medical Device Problem Codes Unable to Obtain Readings (1516); Low Readings (2460)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/10/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "the thermometer wont read his temperature most of the time.Othertimes it takes 20 minutes before he gets a read and read is never going above 96.".
 
Additional Manufacturer Narrative
The customer reported receiving a medline digital oral thermometer that was giving inaccurate readings during use, and reported that the thermometer would not read his temperature most of the time, other times it takes 20 minutes before he gets a read, and the read is never going above 96.Medline received one oral digital stick thermometer in used condition; the sample came with a working battery, was tested according to the instructions for use, the signaling tone was noticeable, and five body temperature readings were between 97.9 and 98.2 f.The complaint was not confirmed, and the cause was not confirmed with the received sample.The sample worked as intended, and medline was unable to confirm the complaint issue or that the product failed to meet specifications.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
THERMOMETER,DIGITAL,ORAL,F/C,30 SEC
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
diana arizmendi
three lakes drive
northfield, IL 60093--275
MDR Report Key50884411
Report Number1417592-2026-50009
Device Sequence Number10707891
Product Code SDV
UDI-Device Identifier00888277116514
UDI-Public00888277116514
Combination Product (Y/N)N
Initial Reporter StateDE
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMDS9950H
Device Lot Number240730001
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/25/2025
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/10/2025
Initial Report FDA Received Date09/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-