Customer reported the nurse anesthetists are complaining about the injection ports failing.Medline received 3 boxes of iv administration set; 15 samples were randomly selected for testing, hooked up to an iv administration bag and observed for leaking or the reported issue of the anti-reflux port failing, and all connections were checked thoroughly.This was not observed in any case, and the complaint was not confirmed.Cause was unconfirmed defect.Not confirmed with received sample; no further action was taken because objectionable trends were not seen for this issue, and division has been informed and will continue to monitor trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|