Medline Industries, LP MEDLINE; IV ADMIN SET, 15DRP, 2 NF PORTS, 108" Back to Search Results
Catalog Number DYNDST1512
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/28/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "the nurse anesthetists are complaining about the injection ports failing".
 
Additional Manufacturer Narrative
Customer reported the nurse anesthetists are complaining about the injection ports failing.Medline received 3 boxes of iv administration set; 15 samples were randomly selected for testing, hooked up to an iv administration bag and observed for leaking or the reported issue of the anti-reflux port failing, and all connections were checked thoroughly.This was not observed in any case, and the complaint was not confirmed.Cause was unconfirmed defect.Not confirmed with received sample; no further action was taken because objectionable trends were not seen for this issue, and division has been informed and will continue to monitor trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
IV ADMIN SET, 15DRP, 2 NF PORTS, 108"
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
sullivan bluhm
three lakes drive
northfield, IL 60093--275
MDR Report Key50884492
Report Number1417592-2026-50019
Device Sequence Number10667626
Product Code FPA
UDI-Device Identifier20884389975951
UDI-Public20884389975951
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K093773
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberDYNDST1512
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer05/08/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/28/2023
Initial Report FDA Received Date09/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-