| Catalog Number |
00515047500 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/12/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during surgery, the unit was not working.After final investigation, battery expulsion was confirmed.There was no patient impact or delay reported.Due diligence is complete as no further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint is recorded by zimmer biomet under (b)(4).Visual inspection of the interior of both packs found the batteries had leaked electrolyte inside of the packs.Additionally, the devices would not power on with the original battery pack, but both did power on and function normally in both modes when connected to a lab battery pack.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.The root cause is attributed to a manufacturing issue exacerbated by a design issue.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor trends.G2: singapore.
|
| |
|