• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Respironics, Inc. BiPAP S/T C-Series; Ventilator, Continuous, Non-Life-Supporting

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Respironics, Inc. BiPAP S/T C-Series; Ventilator, Continuous, Non-Life-Supporting Back to Search Results
Model Number IN1061X
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/26/2026
Type of Reportable Event Malfunction
Event or Problem Description
A device was returned to a third-party service center in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The device was not in patient use.During the evaluation of the device, the third-party service center visually inspected the device and found evidence of foam degradation and dirt and moisture contamination.Additionally, the evaluation of the device data identified unrelated error codes.These error codes are consistent with normal device operation and expected use conditions and are not considered reportable under medical device reporting requirements.No evidence was identified to suggest that these error codes contributed to or caused the reported event.The device has been scrapped.At this time, no further investigation can be performed.If any additional information is received, a follow up report will be filed.
 
Additional Manufacturer Narrative
The affected products with field complaints alleging pe-pur foam degradation have undergone the establishment of complaint codes specifically identifying foam degradation, for dream station cpap/bipap series.A field correction action (2021-06-a) was initiated, and medical device recall letters were issued.A field action was initiated to replace pe-pur foam in affected products; degradation was added to the respective dfmeas and the risk management file that was current at the time ¿ er (b)(4) (nirvana risk management matrix) was updated to include hazards associated with foam degradation based on complaints analyzed through various health hazard evaluations (hhes).Risk tags ¿ degrad04, irrit05 reflected the safety risk assessment of design containing the affected pe-pur foam.These complaints were monitored and trended in accordance with the complaint trending procedures.If the complaints were determined to meet the criteria for escalation, they were escalated for risk assessment through the post market clinical escalation process.As of the date of submission for this report, there is no indication that complaints for the failure in question have led to unacceptable levels of severity or probabilities of harm, and therefore no further action will be pursued.Dhr review was performed for the complaint device sn.Review confirms that the device met the final acceptance criteria and was released for final distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BiPAP S/T C-Series
Common Device Name
Ventilator, Continuous, Non-Life-Supporting
Manufacturer (Section D)
Respironics, Inc.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
Respironics, Inc.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
+1(724)733-0200
MDR Report Key50884604
Report Number2518422-2026-046362
Device Sequence Number21288568
Product Code MNS
UDI-Device Identifier00606959039100
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K102465
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation 505
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberIN1061X
Device Catalogue NumberIN1061X
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/17/2026
Initial Report FDA Received Date09/24/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/19/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1973-2021
Patient Sequence Number1
Patient SexUnknown
-
-