The account reported finding leaking, and the customer also reported backflow with this product.A sample was received for investigation; the device was attached to the port of a primary set with a saline filled syringe, and it was also evaluated by connecting a catheter to the male luer and the saline syringe to the female luer.No leakage was observed at either end of the device, the complaint is not confirmed, and the reported issue could not be confirmed since no leakage was observed.(b)(6)does not have a check valve that would prevent the reported backflow issue; the complaint has been provided to the finished good manufacturing facility for further evaluation and the division will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|