| Model Number |
106531US |
| Medical Device Problem Code |
Failure to Charge (1085)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/03/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the patient had been having emergency backup battery (ebb) faults/ replace backup battery alarms since the night before.The ebb was replaced, but yellow wrench alarms/ ebb faults were still occurring.The log captured ebb faults alarms starting from (b)(6) 2026 at 2304 and continued for the remainder of the log files.The ebb fault was due to the ebb failing the load test.It was intended to exchange the system controller.
|
| |
|
Additional Manufacturer Narrative
|
|
No further information was provided.A supplemental report will be reported once the manufacturer¿s investigation is completed.
|
| |
|