The customer states the meter is providing consistent out of ranges and confirmed solution is not expired.Medline received 1ea evencare g2 blood glucose meter in used condition; test strips and control solutions used by the user were not sent for the investigation, and testing was performed according to the instructions for use using control test strips and control solutions.Results using the control strips and solutions were 55 and 196mg/dl for low and high solutions respectively, which were within control solutions range of 41-71 mg/dl and 167-227 mg/dl; the sample turned on/off and control solution results were within testing range.The complaint was not confirmed, no root cause was identified, and the division will continue to monitor trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|