| Model Number |
HexaVue SCC 1.0 |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/13/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The user facility reported that during a patient procedure the monitors connected to their hexavue custom room rack ¿went black and then came back on¿.The patient procedure was completed successfully.No report of injury.
|
| |
|
Additional Manufacturer Narrative
|
|
A steris service technician arrived onsite following the reported event to inspect the hexavue custom room rack.During the technician¿s inspection, he was unable to duplicate the reported event and found no issues with the function or operation of the hexavue custom room rack or monitors; the devices were operating to specification.The technician returned the hexavue custom room rack to service, and no additional issues have been reported.
|
| |
|