• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; DBD-MONITOR, BP, AUTO, DIGITAL, ADULT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; DBD-MONITOR, BP, AUTO, DIGITAL, ADULT Back to Search Results
Catalog Number MDS1001
Medical Device Problem Codes Unable to Obtain Readings (1516); Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/13/2026
Type of Reportable Event Other
Event or Problem Description
It was reported that "digital blood pressure is reading up and down and not giving a reading accuracy.Tried to leave on longer and adjusted the cuff but still not working.Does not show no error code.".
 
Additional Manufacturer Narrative
The customer reported receiving a medline bp monitor that is inaccurate and reported that the blood pressure cuff would inflate and did not stop, no error code was displayed, and no injury occurred or medical intervention was required.At this time, a sample(s) was not received for investigation; investigation included review of recent complaint trends/scar activity for the last three trend reports, and the vendor stated, "after checking the process of our company, we found no problems related to the process leading to complaints, but our company will record and observe the trend." the complaint was not confirmed due to the fact that this issue can only be confirmed by further investigating the sample or a provided photo; the listed cause was "sample required to determine cause." no trends were identified for the specified material number and its corresponding defect type, the division will continue to monitor trends, and there has been no change to the reportability decision associated with this complaint.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
DBD-MONITOR, BP, AUTO, DIGITAL, ADULT
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
vamshi thupakula
three lakes drive
northfield, IL 60093--275
MDR Report Key50884940
Report Number1417592-2026-50099
Device Sequence Number10707915
Product Code DXN
UDI-Device Identifier00888277629656
UDI-Public00888277629656
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K193627
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Device Catalogue NumberMDS1001
Device Lot NumberL200580005
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/13/2026
Initial Report FDA Received Date09/24/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age78 YR
Patient SexMale
Patient Weight220 LB
-
-