The account reported finding open perforation pouch, and the complaint transcript states the issue did not occur during use and no injury occurred.Photos were provided for evaluation; finished good production records, finished good pack build, and trending were reviewed.The photos demonstrate the clear perf bags visibly open on the seam of the two sides of the bag, confirming the reported issue.Based on the complaint details, the pouch was damaged on an outside source at some point during production, transit, or storage; although the reported issue was confirmed, medline was unable to determine at what point the damage occurred, and the division will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|