The surgical issue questionnaire was reviewed for potential causes of the reported issue.Based on this review, implanting the device after the expiration date, not prepping the surface of the skin with an antiseptic solution, and multiple tunneling attempts have been ruled out.Additionally, the patient touching or picking at the wound has been ruled out.However, the patient is a poor surgical risk as they are prone to infections.Per the freedom peripheral nerve stimulator system implantation of neurostimulator instructions for use, 05-20200 rev.8 "poor surgical risks - peripheral nerve stimulators should not be used on patients who are poor surgical risks or patients with multiple illnesses, active general infections, or other potential surgical risks as identified by the clinician." additionally, non-compliance to the ifu was identified as the surgical site was closed using staples."closures are usually done in two steps, with absorbable sutures for the deeper layer and either nonabsorbable or absorbable sutures for the superficial layer."
a curonix representative conducted a review of sterilization and packaging records for the respective product lot; curonix has confirmed that the product was delivered sterile, validated sterilization parameters were used, and sterile barriers were verified to be intact following packaging.The stimulator is used to treat pain.The cause of the infection is attributed to two identified root causes: patient is a poor candidate due to health, weight, age, or mental capacity as they are prone to infections (user error-clinician.) non-compliance to the ifu as the surgical site was closed using staples (user error-clinician).Rates reviewed at the most recent complaint trending meeting do not indicate a significant increase in surgical issues.Surgical issue rates remain acceptably low; thus, a capa is not required at this time.Surgical issue rates will continue to be tracked and trended.
|