| Model Number |
M312 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/25/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this insertable cardiac monitor (icm) device was interrogated and showed end of service.Technical services recommended device replacement and requested the device be returned for analysis.Additional information indicated that the patient had been wearing a magnet over the implant area since the time of implant.The potential effects were discussed; patient education and physician notification were recommended.No adverse patient effects were reported.This icm remains in service.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.
|
| |
|