It was reported that during a da vinci-assisted cholecystectomy procedure, a minor burn injury was observed on the inside of the patient¿s abdomen at the trocar site where the permanent cautery hook (pch) instrument was used.The injury was assessed as a first-degree burn with a black, charred appearance.No bleeding was observed, and the surgeon did not provide additional treatment.The procedure was successfully completed robotically.The patient had no postoperative complications, and there is no concern for long-term complications related to this event.The patient was discharged the following day without further issues.The surgeon was unable to determine the cause and reported no observed evidence of arcing during the procedure, but noted a potential contribution from the pch instrument, green cable, or erbe generator.No damage or abnormalities were identified on the instruments or accessories.An erbe vio dv generator was used with settings of 3 for monopolar cut/coag and bipolar coag.The grounding pad was inspected prior to use; no defects or abnormalities were identified, and it was properly placed on the patient¿s left thigh.
|
|
Intuitive surgical, inc.(isi) did not receive the permanent cautery hook (pch) instrument to perform failure analysis.A review of the customer-provided image shows that the tissue does appear to be desiccated; however, the source or cause of the burn could not be determined.
|