| Model Number |
OTV-S7PROH-HD-L08E |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the device manufacturing date.\
\
updated fields: h2, h4, h11.\
\
should additional relevant information become available, a supplemental report will be submitted.\
\
olympus will continue to monitor field performance for this device.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of event is unknown.Additional information will be provided if available.\
\
the final investigation has been completed.\
\
based on the final investigation of the device, failure identified during the inspection have been confirmed.\
\
a root cause could not be identified.Based on the results of the investigation, the most probable cause of the malfunction was traced to component failure.\
\
should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
The customer returned the camera head, which is an olympus asset with no reported complaint.During the initial evaluation of the returned device, a failed leak test was observed.There was no patient involvement.
|
| |
|
Event or Problem Description
|
|
No additional information received from the customer.
|
| |
|