Model Number 72404302
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 02/01/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient with this inflatable penile prosthesis (ipp) developed an infection approximately ten months after implantation.The physician treated the infection with antibiotics.A surgical procedure was performed in which the existing ipp was explanted and replaced with a new inflatable penile prosthesis (ipp).The physician indicated that the infection was attributed to a microbial source and not to foreign material associated with the device.There were no further patient complications reported.
 
Additional Manufacturer Narrative
Upon receipt at our quality assurance laboratory, this inflatable penile prosthesis (ipp) underwent a thorough analysis.The pump was microscopically inspected, functionally and leak tested.Under microscope observation, contamination (bodily fluid) was found within the ms pump.As a result, the functional test was not performed.The pump passed the leak test.The cylinders were microscopically inspected and leak tested.The first cylinder exhibited a hole in the proximal end of the cylinder from sharp instrument.Both cylinders had wear at fold in the middle of the cylinder.The first cylinder did not pass the leak test, however, the second one did.The reservoir was microscopically inspected, and leak tested.No anomalies were found with the reservoir.The device history record (dhr) review confirmed that the device met all material, assembly and performance specifications prior to distribution.A review of the device instructions for use (ifu) was completed and did not reveal any evidence of device misuse, off label use, or failure to follow instructions.A risk review was completed and confirmed that the event of infection was defined in the product risk documentation.This event type has been accounted for during analysis to support acceptable risk benefit for the product.Based on this investigation, the investigation conclusion code of adverse event related to patient condition was chosen as the physician indicated that the infection was attributed to a microbial source and not to foreign material associated with the device.The allegation of infection is addressed in our labeling and risk documentation.This record was identified during a retrospective review of complaint events; an electronic medical device report (emdr) was not previously submitted per applicable reporting criteria.Boston scientific has opened a nonconforming events and prevention investigation ((b)(4)) to determine the potential cause for why an emdr was not previously submitted to the food and drug administration (fda) in accordance with boston scientific's established medical device reporting process and fda adverse event reporting requirements.Information gained through these investigation efforts will be utilized to take action(s), as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS 700 LGX MS Pump
Common Device Name
prosthesis, penis, inflatable
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key50885682
Report Number2124215-2026-50940
Device Sequence Number16140987
Product Code JCW
UDI-Device Identifier00878953005737
UDI-Public(01)00878953005737(17)290408(10)1100482728
Combination Product (Y/N)N
Initial Reporter CountryGR
PMA/510(K) Number
N970012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number72404302
Device Catalogue Number72404302
Device Lot Number1100482728
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer03/26/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/26/2026
Initial Report FDA Received Date09/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/09/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;