| Catalog Number |
05964067001 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
We received an allegation about discrepant results for 1 patient sample tested with glucose hk gen.3 assay on a cobas 8000 cobas c 502 module.Initial result: 205 mg/dl.Multiple abnormal aspiration alarms prompted a repeat of the sample.No questionable result was reported outside the laboratory.Repeat result: 104 mg/dl.The repeat result was reported outside the laboratory.
|
| |
|
Additional Manufacturer Narrative
|
|
The glucose hk gen.3 reagent lot number was 953833.The expiration date was not provided.A general reagent issue can be excluded, as no further issues were reported and a correct rerun was performed with the same reagent.The customer observed several sample short and clot alarms in parallel.The field service engineer (fse) replaced the liquid level detection circuit board (lld pcb) and the sample probe.He then verified that the test and the instrument were performing within specifications.A pcb issue can be excluded, as it was used for multiple years in a high-throughput environment and its overall consumption at a low level was stable.The service actions performed by the fse resolved the issue in a commonly trained troubleshooting process.The cause of the event was due to a component failure.
|
| |
|