| Model Number |
KR2110X15B |
| Medical Device Problem Code |
Failure to Charge (1085)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/31/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The manufacturer received information alleging that the ventilator displayed a service required alarm and experienced poor charging.There was no serious patient harm or injury that occurred or was reported.The trilogy evo was returned and evaluated by philips engineer.The device evaluation confirmed that when connected to alternating current (ac) power, the power led light does not turn on, and the battery did not charge.Replacement of the power supply was required to address the condition.In addition, the event logs were reviewed and confirmed the service required alarm for error code 201-evt_power_buzzer_fail, indicating possible no sound from buzzer at start and replacement of the buzzer was required.Following completion of the required service activities, the device passed final testing.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported failure of 'poor battery charging' is accommodated in the product risk management file as being linked to hazard with description loss of function/loss of primary function.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.Dhr review was performed for the complaint device sn (b)(6).Review confirms that the device met the final acceptance criteria and was released for final distribution.
|
| |
|
Search Alerts/Recalls
|
|
|