• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. Inceptiv; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. Inceptiv; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR Back to Search Results
Model Number 977119
Medical Device Problem Codes Failure to Deliver Energy (1211); Energy Output Problem (1431)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/22/2026
Type of Reportable Event Serious Injury
Event or Problem Description
Information was received from a patient's representative regarding an implantable neurostimulator (ins).The caller reported the patient had their therapy off because they had to go in the hospital for unrelated concerns and didn't want the therapy interfering with anything.Caller said they were trying to put the therapy back on today, but they couldn't get the patient to feel the "current." caller said patient was typically on group c with intensity set to 7, which was what they were at when the therapy was turned off.Caller stated that because the patient wasn't feeling the current, they played with the settings and lowered the amount to 5.4 to see if that would work.When they tried to increase the amount, they got an error message that said 'contact your clinician.Your therapy turned off automatically.Service code: 323.' agent reviewed information and redirected to healthcare provider to further address the issue.Agent reviewed their role and a manufacturer representative (rep).Agent explained the issue would need to be resolv ed in person.Agent advised caller to reach out to healthcare provider (hcp) directly to page a rep if they wanted to coordinate with a local rep faster.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Inceptiv
Common Device Name
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
+1(763)526-5677
MDR Report Key50886095
Report Number3004209178-2026-16259
Device Sequence Number9827151
Product Code LGW
UDI-Device Identifier00763000984687
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P840001
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number977119
Device Catalogue Number977119
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/22/2026
Initial Report FDA Received Date09/24/2026
Date Device Manufactured11/18/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age84 YR
Patient SexFemale
-
-