| Model Number |
KR2110X15B |
| Medical Device Problem Codes |
Electrical /Electronic Property Problem (1198); Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The manufacturer received information that a trilogy evo ventilator was returned to the manufacturer service center for evaluation.There was no patient involvement, and no harm or injury reported.On evaluation, ventilator service required event was noted in the event log indicating failure of the internal lithium-ion battery and required replacement.After battery replacement the device passed final tesing.Additional findings not related to the malfunction/issue: dust contamination was noted on the machin flow sensor; the machine flow sensor was replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Dhr review was performed for the complaint device sn (b)(6).Review confirms that the device met the final acceptance criteria and was released for final distribution.The reported failure of the trilogy evo internal battery is accommodated in the product risk management file as being linked to hazard with description loss of function/loss of primary function.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.
|
| |
|
Search Alerts/Recalls
|