| Model Number |
WA50052A |
| Medical Device Problem Codes |
Poor Quality Image (1408); Positioning Problem (3009)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/23/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the video telescope produced a good image when viewed up close; however, when panned out, the image quality deteriorated.In addition, the angle adjustment on the main body twisted to the right side and was reported to be very loose.The issue occurred during a therapeutic procedure was delayed for less than 30 minutes under sedation and was completed using an olympus backup device.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device was returned to olympus for inspection and the reported failure was confirmed.A definitive root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: that the image was unstable due to loose adapter.In addition, the following reportable malfunctions were identified during the device evaluation: image was out of field due to outer tube being bent and dust particles under distal cover glass.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|