SHIRAKAWA OLYMPUS CO., LTD. ULTRASONIC GASTROVIDEOSCOPE Back to Search Results
Model Number GF-UE160-AL5
Medical Device Problem Codes Break (1069); Detachment of Device or Device Component (2907); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/26/2026
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported that the ultrasonic gastrovideoscope had a damaged transducer during sedation for the procedure.During the initial evaluation of the device, missing thixotropic adhesive (ke45) and a cut in the pink probe were observed.There was no patient injury reported.
 
Additional Manufacturer Narrative
The final investigation has been completed.The failure reported by the customer and identified during the inspection have been confirmed.Based on the results of the investigation, the most probable cause of the malfunction reported by the customer and identified during the inspection was traced to component failure, without any design or manufacturing issue.A review of the service history record found no deviations that could have caused or contributed to the reported issue.Olympus will continue to monitor the field performance of this device and evaluate any additional information received through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASONIC GASTROVIDEOSCOPE
Common Device Name
ULTRASONIC GASTROVIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
+1(508)207-7661
MDR Report Key50886519
Report Number3002808148-2026-33691
Device Sequence Number10744807
Product Code ODG
UDI-Device Identifier04953170356261
UDI-Public04953170356261
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K051541
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberGF-UE160-AL5
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/28/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/26/2026
Initial Report FDA Received Date09/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/09/2005
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-