• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Curonix LLC Freedom Peripheral Nerve Stimulator

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Curonix LLC Freedom Peripheral Nerve Stimulator Back to Search Results
Model Number STQ4-RCV-A0
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Swelling/ Edema (4577)
Date of Event 09/04/2026
Type of Reportable Event Serious Injury
Event or Problem Description
The patient was taken to the hospital on (b)(6) 2026 due to swelling in the lower extremity and concerns of infection.Antibiotics were prescribed the week prior after a follow up appointment with the implanting clinician.As of (b)(6) 2026, the patient completed their antibiotics, they are doing well, and resumed using the neurostimulator without issue.
 
Additional Manufacturer Narrative
The other adverse events issues issue questionnaire was reviewed for causes of the reported issue.Based on this review, implanting the device off-label, implanting the device too close to the targeted nerve, and inadvertently puncturing bone or tissue during the procedure have been ruled out.Additionally, the patient having pneumonia and dehydration have been ruled out.A curonix representative conducted a review of sterilization and packaging records for the respective product lot; curonix has confirmed that the product was delivered sterile, validated sterilization parameters were used, and sterile barriers were verified to be intact following packaging.The stimulator is used to treat pain.The cause of the swelling is unknown.The investigation findings do not lead to a clear conclusion about the cause of the reported issue.Therefore, conclusion has been selected as unable to determine root cause.Rates reviewed at the most recent complaint trending meeting do not indicate a significant increase in other adverse events issues.Other adverse events issues rates remain acceptably low; thus, a capa is not required.Other adverse events issues rates will continue to be tracked and trended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Freedom Peripheral Nerve Stimulator
Common Device Name
Peripheral Nerve Stimulator
Manufacturer (Section D)
Curonix LLC
1310 park central boulevard s.
pompano beach FL 33064
Manufacturer (Section G)
Curonix LLC
1310 park central boulevard s.
pompano beach FL 33064
Manufacturer Contact
alicia pagliaro
1310 park central boulevard s.
pompano beach, FL 33064
+1(800)965-5134
MDR Report Key50886588
Report Number3010676138-2026-00200
Device Sequence Number15900557
Product Code GZF
UDI-Device Identifier00850051034406
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K233162
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberSTQ4-RCV-A0
Device Lot NumberSWO260729
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/08/2026
Initial Report FDA Received Date09/24/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/14/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
Patient SexMale
-
-