| Model Number |
107801 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Bacterial Infection (1735); Aortic Valve Insufficiency/ Regurgitation (4450); Skin Infection (4544)
|
| Date of Event |
08/31/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient had a history of short-to-shield, which was still present after an external driveline repair was performed on (b)(6) 2026 (manufacturer report 2916596-2026-02141).The patient had come from another implanting facility and was a \"transfer and legacy encounter\".A heartmate 2 (hm2) to heartmate 2 pump only exchange was completed for short-to-shield on 09sep2026.The patient was in stable condition.A pre-operative transesophageal echocardiogram (tee) showed moderate to severe aortic insufficiency (ai).A post-operative echocardiogram showed severe ai which the site was addressing.It was also reported that the patient had chronic a chronic driveline exit site infection (dlesi) and was on doxycycline.The driveline exit site (dles) had moved to the left side with the new heartmate 2 pump implant.The patient was put on suppressive antibiotic therapy and managed by infectious disease physicians.The event had not resolved as the patient's condition was considered to be chronic.Images were not available.
|
| |
|
Additional Manufacturer Narrative
|
|
No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is completed.
|
| |
|