It was reporting that after pump preparation was completed, the heartmate touch (hmt) abruptly lost connection to the system controller.All connections were checked and secured.The bluetooth connection was stable.Extended silence was noted to be disengaged.There were attempts to reconnect to the controller, but the screen would not verify the connection.Images were submitted of the issue.A new hmt was obtained and there were no further issues at that time.During the case, multiple attempts were made to start the pump, but the command would not go through and the pump would not start.The power was stable at 0.8w.Usual alarms were present on the controller, including pump off alarms, low flow alarms, and backup battery not installed alarms.A new ventricular assist device (vad) cart was obtained with a hmt, a bluetooth adapter, and a new system controller.A controller exchange was performed.No further connectivity issues or alarms occurred after the exchange.The customer requested a replacement controller.
|