| Model Number |
106762 |
| Medical Device Problem Codes |
Premature Discharge of Battery (1057); Disconnection (1171); Electrical Power Problem (2925)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/14/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the patient had multiple low voltage advisory alarms and stated that their batteries were not draining equally.The battery connected to the white cable on the system controller was draining much faster.The patient cleaned all peripheral equipment and changed their batteries and battery clips, but the alarms did not resolve.The alarms continued to occur on all batteries and different sets of battery clips.Log files were submitted.Upon review, the heartmate 2 log files recorded 228 low battery advisory events on 14sep2026 from 2:27:13pm to 4:32 pm.The low battery hazard advisory events were recorded at times when the white power lead was disconnected.A root cause for the event was not able to be determined based on the log files.Follow-up from the customer reported that the patient's primary pocket controller was successfully exchanged and the alarms had resolved.The patient would continue to be monitored.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d4: device expiration date cannot be determined.The udi number in d4 is reflective of all available information.No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is completed.
|
| |
|