| Catalog Number |
DYNJLHH2 |
| Medical Device Problem Code |
Unsealed Device Packaging (1444)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This vmsr report summarizes 1 malfunction event involving dynjlhh2.The account reported that the item was not sealed correctly and therefore was not sterile.Evaluation of five returned samples confirmed that the items were not properly sealed.Based on the complaint details, sample evaluation, and investigation, the reported issue was confirmed.No patient information was available.No patient or user injuries were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This vmsr submission is part of a retrospective data remediation review.The included event represents 1 malfunction event.The account reported that item dynjlhh2 was not sealed correctly and therefore was not sterile.The reported issue involved lot numbers 06924030011 and 06924050012.Five samples were returned for evaluation.Medline quality performed a visual inspection of the returned samples and confirmed that the items were not properly sealed.Finished goods production records were reviewed as part of the investigation.Based on the complaint details, sample evaluation, and investigation, the reported issue was confirmed.The vendor was notified of the issue for further evaluation, and the supplier quality team will continue to monitor this issue for trending purposes.Of the events summarized, 5 samples were returned for evaluation.Information regarding reprocessing or reuse was not identified during the investigation.
|
| |
|