The customer reported issue of product being crystallized and hardened with item mnk0015 dressing,honey,therahoney,1.5 oz tube.Medline reviewed the sample returned and was able to confirm an issue with a peanut butter color and the tubes bursting.The issue was confirmed with the samples.Cause was documented as mfg: machine/automation process error under dc: supplier error non-asia(medline brnd); the supplier was notified, the qa team and the division are working to assess appropriate corrective actions with the vendor, and the division will continue to monitor trending through the quarterly trend reports.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|